The method of analysis adopted follows the indications of ISO 140 part
2, and more generally of ISO 5725-1986, for a uniform level experiment.
In this I.L.T. the levels are the 1/3 octave frequency bands from 100 to
5000 Hz.
Each combination of a laboratory and a level is called a cell of the
precision experiment. Ideally in each cell of this I.L.T. there should
be 5 replicate results.
It was decided to calculate the repeatability r and the reproducibility
R only among the 21 laboratories in which the mounting of the specimens
had been overseen by the I.L.T. supervisor.
Regarding the remaining three laboratories, their repeatability and
reproducibility values with respect to the other laboratories shall be
verified.
2) Preliminary considerations
The following tables must be filled in, as indicated by ISO 5725:
- The collation of the original data;
- The measures of spread for each cell of the previous
tables. The formula used for computing the cell spread sij
is:
where yijk is one of the results
of that cell.
- The cell averages. The formula used for calculating
the cell averages is:
3) Statistical tests for outliers
Outliers are entries that deviate so much from the comparable numbers
in a table that they can be considered as irreconcilable with the other
data.
According to ISO 5725-1986, the following statistical tests were performed:
4) Repeatability values r
The repeatability value r is the value below which the absolute
difference between two single results obtained under repeatability conditions
may be expected to lie with a probability of 95 %.
Assuming that the distribution of the random errors occurring in every
single test result is approximately normal, the repeatability r can
be calculated as follows:
where sr2 is the repeatability variance, that is, the average of the within-laboratory variances si2 , computed as follows (for convenience, the subscript j which indicates the level is omitted):
Regarding the repeatability of single numbers Rw, it has been computed among the single number quantities related to each series of results.
5) Reproducibility values
The reproducibility value R is defined as the value below which
the absolute difference between two single results obtained under reproducibility
conditions may be expected to lie with a probability of 95 %.
Assuming that the distribution of the random errors occurring in every
single test result is approximately normal, the reproducibility R can be
calculated as follows:
where sr2 is the reproducibility variance and sL2 is the between-laboratory variance calculated as follows (for convenience, the subscript j which indicates the level is omitted):
Concerning the reproducibility of single numbers Rw, it has been calculated using the single numbers quantities related to the within-laboratory averages.
6) Verification of repeatability and reproducibility values
According to ISO 140 part 2, it is possible to verify that the repeatability and reproducibility values just calculated actually correspond to a 95% probability, as required by the definition, by means of the data from which they have been computed.
It must be noted that the repeatability and reproducibility values for single number quantities are computed on integer numbers determined by the procedure indicated in ISO 717-1, consequently it is expected that in this case the basic model will not be very suitable.
The correct critical difference to be used for each frequency band is (where n is the number of determinations):
For n=1, this reduces to the reproducibility value R as expected.
The critical differences thus calculated for each frequency band are
compared with the observed differences of the averages in that frequency
band. The overall number of comparisons is 3780 (coincidentally the same
as for the repeatability verification).
It is expected that the critical differences be exceeded in no more
than 5% of the cases, i.e. no more than about 190 times.
7) Verification of excluded laboratories
According to ISO 140 part 2, the three laboratories which were excluded from the calculation of the repeatability and reproducibility values may verify the proper operation of their own test procedure by using the I.L.T. results.
where sx is the standard deviation of the 5 results and r is the repeatability value determined in I.L.T.
The appropriate critical difference for this case is:
These critical differences should not be exceeded in more than 5% of the cases, i.e. not more than twice.